ALLSHENG ALLSHENG
Two Methods to Measure Protein Concentration by Fluorescence
ALLSHENG Release time:2022-09-14 ALLSHENG Source: ALLSHENG Pageviews:4424

Protein quantification is an essential part of many protein biology workflows and is a necessary step prior to commonly used techniques such as protein electrophoresis, western blotting, mass spectrum, and immunoassays. Compared with traditional colorimetric methods such as BCA and Bradford, the fluorescence method has a wider dynamic range and higher accuracy.


Fluorescent protein assay kits are highly selective for proteins and can also be highly accurate in the presence of reducing reagents. The kit can not only be tested on a general-purpose fluorometer, but also can be tested on a microplate reader with fluorescence function.


Reagents and Instruments

•Qubit™ Protein Assay Kits (Thermo: Q33212)

•Qubit 4 Fluorometer (Thermo)

•Fluo-200 Fluorometer (ALLSHENG)

•Feyond-A300 Multi-mode Microplate Reader (ALLSHENG)


Method

•Standard Preparation

Sample volume

10μL

Standard preparation

The highest concentration of 400ng/uL protein standard in the kit was diluted to prepare 6 concentrations of 300, 200, 100, 50, 25, 12.5ng/uL, and the kit concentration ranged from 12.5 µg/mL to 5 mg/mL.


•Steps

(1) According to the reagent instructions, prepare dye Mix working solution and 0, 200 and 400ng/uL calibration point reagents. After the calibration curve is completed, detect on Qubit4 and Fluo-200 respectively.

(2) After adding 10uL of standard solution to each well of a 96-well black flat-bottom microplate, add 190uL of Mix working solution to each well, mix well and protect from light for 5 minutes, then start the instrument detection, the standard and blank layout are shown in Figure 1, the instrument parameters are shown in Table 1.


Table 1 Feyond-A300 Parameter Setting

Reagent wavelength

EX/EM: 470/525

PMT gain

Medium low

Stabilization time

150ms

Integration time

40μs


Figure 1 Sample Layout Interface

微信图片_20220906091152.png


Result

(1) Using the built-in data processing software of Feyond-A300, the standard curve fitting of the protein standard concentration and the fluorescence intensity, and the four-parameter (4PL) fitting method was used to obtain the curve equation and the curve fitting degree, and R2 was close to 1 (Fig. 2).


Figure 2. Protein Standard Curve

微信图片_20220906091316.png


(2) When the concentration range is 12.5 ng/μL~400 ng/μL, the detection concentration deviation of Fluo-200 and Qubit4 is within 10%, which is a reasonable deviation of Qubit. Except for the lowest point close to the limit concentration, the detected concentration of Feyond-A300 was also within a reasonable deviation compared with Qubit4, and was closer to the theoretical concentration (Table 2).


Table 2 Test Data of the Three Instruments

微信图片_20220906091321-en.jpg


Summary

Qubit fluorescence detection protein kit is not only applicable to Qubit instruments, but also applicable to the fluorescence detection function of Feyond-A300, and the accuracy is basically the same. The fluorometer detection does not need to establish a standard curve, only needs to calibrate the standard points, and the concentration can be directly tested, which is simple and convenient. Feyond-A300 standard curve storage function can also meet this requirement. After establishing a standard curve once, save the standard curve in the standard curve library, and then only need to retrieve the standard curve to directly give the concentration of the unknown sample.    

图片1-en.jpg

微信图片_20220906091326.png

Contact us
ALLSHENG
ALLSHENG
Success